Voyager DP-263-S Podręcznik Użytkownika Strona 281

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 280
Impact of Changing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-51
5
Figure 5-12 Effect of Vertical Scale in Digitizer Settings
Suggested settings
Use the settings listed below as a starting point for Vertical
Scale. Adjust as needed to bring the signal into the proper
range.
Vertical
Scale
Decrease if
signal is weak
Increase if
signal exceeds scale
Mass Range (Da) Vertical Scale
0–10,000 1,000 mV
10,000–50,000 200 mV
>50,000 50 mV*
* Not available on the Signatec digitizer.
Przeglądanie stron 280
1 2 ... 276 277 278 279 280 281 282 283 284 285 286 ... 611 612

Komentarze do niniejszej Instrukcji

Brak uwag