Voyager DP-263-S Podręcznik Użytkownika Strona 382

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 381
Chapter 6 Acquiring Mass Spectra
6-70 PerSeptive Biosystems
6
Figure 6-22 Sequence Run List
Hint: You can show and hide columns by selecting a
column, then selecting Show/Hide from the
View menu.
2. Click the scroll bar at the bottom of the grid to display
columns in the grid that are not in view.
3. Click on a cell to activate it.
4. Enter Run List parameters as described in Table 6-4.
Run list
Table 6-4 Run List Parameters
Field Description
Sample Position
(required entry)
Position from which to acquire data.
Data File
(required entry)
Base name of the data file to acquire. During acquisition, the
software appends a 4-digit suffix starting at 0001 to the base
name. For example, if you specify a base name of SAMPLE, the
final data file name is SAMPLE0001.DAT.
Przeglądanie stron 381
1 2 ... 377 378 379 380 381 382 383 384 385 386 387 ... 611 612

Komentarze do niniejszej Instrukcji

Brak uwag