Chapter 5 Optimizing Instrument Settings
5-72 PerSeptive Biosystems
5
5.4.3.5 Optimizing Grid Voltage%
Overview
You can use this procedure as an alternative to the procedure
in Section 5.4.3.4, Optimizing Delay Time. However,
optimizing Delay Time is the recommended procedure.
Optimizing the Delay Time ensures maximum stability of the
high voltage power supplies, and therefore results in better
mass accuracy.
You can use this procedure when operating in Linear mode or
Reflector mode. Start with a standard .BIC file, optimize the
Grid Voltage%, and leave the Delay Time unchanged.
NOTE: The Grid Voltage% setting in Reflector mode is
typically significantly lower than the setting in Linear mode
for the same mass range. For example, if Grid Voltage% in
Linear mode is 94%, Grid Voltage% in Reflector mode may
be 70%. (These values are used as an example they may
not be optimal for your system.)
For more
information
For more information, see Section 5.3.2, Understanding Grid
Voltage%.