Chapter 6 Acquiring Mass Spectra
6-82 PerSeptive Biosystems
6
6.7.7 Automatic Calibration
During a Sequence Run
This section includes:
Calibration options in a sequence
Calibration standard requirements
Performing close external calibration
Internal standard calibration considerations
6.7.7.1 Calibration Options in a Sequence
The Sequence Control Panel allows three types of automatic
calibration:
For the mass accuracy provided by each calibration type, see
“Types of calibration” on page 6-7.
Type Function
External Applies calibration constants in a specified .CAL file.
Internal Applies calibration constants if a .CAL file specified
Generates new calibration constants using standard reference
masses specified by a .SET file.
Applies the constants.
Internal-
Update
Applies calibration constants if a .CAL file specified
Generates new calibration constants using standard reference
masses specified by a .SET file.
Applies the constants.
Updates the .CAL file specified, or creates a new .CAL file if
the specified file does not exist.
Use this option to perform a close external calibration, described in
Section 6.7.7.3, Performing Close External Calibration.