Chapter 6 Acquiring Mass Spectra
6-82 PerSeptive Biosystems
6
6.7.7 Automatic Calibration
During a Sequence Run
This section includes:
• Calibration options in a sequence
• Calibration standard requirements
• Performing close external calibration
• Internal standard calibration considerations
6.7.7.1 Calibration Options in a Sequence
The Sequence Control Panel allows three types of automatic
calibration:
For the mass accuracy provided by each calibration type, see
“Types of calibration” on page 6-7.
Type Function
External Applies calibration constants in a specified .CAL file.
Internal • Applies calibration constants if a .CAL file specified
• Generates new calibration constants using standard reference
masses specified by a .SET file.
• Applies the constants.
Internal-
Update
• Applies calibration constants if a .CAL file specified
• Generates new calibration constants using standard reference
masses specified by a .SET file.
• Applies the constants.
• Updates the .CAL file specified, or creates a new .CAL file if
the specified file does not exist.
Use this option to perform a close external calibration, described in
Section 6.7.7.3, Performing Close External Calibration.