Optimizing Instrument Settings Parameters
Voyager
Biospectrometry
Workstation User’s Guide 5-61
5
5.4.3 Optimizing Resolution
This section includes:
•Overview
Acceptable resolution in Delayed Extraction Mode
Optimizing Guide Wire Voltage%
Optimizing Delay Time
Optimizing Grid Voltage%
For information on measuring resolution, see Section 6.5.2,
Calculating Mass Resolution, and
Data Explorer Software
User’s Guide
, Section 6.4, Using the
Mass Resolution Calculator.
Manually
accumulating
spectra to
improve
resolution
The manual accumulation feature of the Voyager system
allows you to acquire a spectrum, examine the spectrum for
signal quality (visually, or using the Resolution or
Signal-to-Noise Ratio calculator), and then accumulate the
spectrum with other previously examined spectra, or discard
the spectrum. This process of accumulating only acceptable
spectra allows you to improve the resolution of the final
spectrum you save. For more information on manually
accumulating spectra, see Section 6.2.2, Manually
Accumulating Spectra from Multiple Acquisitions.