Checking System Status
Voyager
Biospectrometry
Workstation User’s Guide 2-51
2
Table 2-2 System Status Parameters
Parameter Description
System Status
Instrument
State
Possible states are:
ON—Instrument is initialized and high voltage is on.
OFF—Instrument is not initialized and high voltage is off.
FAULT—Indicates a fault condition. Follow the instructions
displayed to correct the fault.
High Voltage Possible states are:
RAMPING—Voltage is turning on.
ON—High voltage is on.
OFF—High voltage is off.
FAULT—Indicates a fault condition. Follow the instructions
displayed to correct the fault.
Source
Chamber
(BA1)
Pressure
Displays the source chamber pressure. Operating pressure ranges
are:
Voyager-DE, Voyager-DE PRO—Less than 10
-6
Voyager-DE STR—Less than 5 x 10
-7
Mirror
Chamber
(BA2)
Pressure
Displays the mirror chamber pressure. Mirror chamber pressure
ranges are:
Voyager-DE PRO—Less than 5 x 10
-7
Voyager-DE STR—Less than 5 x 10
-8
NOTE: This status field is not displayed on Voyager-DE systems.
Acquisition Displays acquisition status:
ON—Instrument is currently acquiring a spectrum.
OFF—Instrument is not acquiring.
Data Storage Displays data storage status:
ON—Software is currently storing a data file.
OFF—Software is currently not storing a data file.