Voyager DP-263-S Podręcznik Użytkownika Strona 334

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 333
Chapter 6 Acquiring Mass Spectra
6-22 PerSeptive Biosystems
6
Figure 6-8 Example of Poor Mass Spectrum for Angiotensin
Figure 6-9 Example of Good Mass Spectrum for Angiotensin
High laser power causing the following:
Matrix peaks and sample saturated
Broad peak
Poor resolution
Mass is
accurate
Minor components
or fragments of major
components ionized
Signal is
saturated
Baseline noise
Peak not saturated
Sharp, narrow peak
resolved to isotopes
Matrix peaks minimized
No minor
components
or fragments
Lower laser power producing:
Przeglądanie stron 333
1 2 ... 329 330 331 332 333 334 335 336 337 338 339 ... 611 612

Komentarze do niniejszej Instrukcji

Brak uwag