Voyager DP-263-S Podręcznik Użytkownika Strona 171

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 170
Loading Sample Plates in the Mass Spectrometer
Voyager
Biospectrometry
Workstation User’s Guide 3-41
3
3.4.1 Ejecting the Sample Holder
In the Voyager Instrument Control Panel, select Eject from the
Sample Plate menu. The Load/Eject dialog box is displayed.
.
Figure 3-4 Load/Eject Sample Plate Dialog Box
The following also occurs:
A Load Status dialog box is displayed during the ejection
sequence which displays hardware status.
High voltage is turned off.
The video monitor displays the sample plate moving.
The sample holder moves out of the main source
chamber, out of the sample loading chamber, and is
ejected from of the instrument.
NOTE: Do not leave the sample holder in the Eject position
for more than a few seconds. Minimize the time the
instrument is exposed to the atmosphere to reduce the time
needed to reach high vacuum when you insert a new plate.
Przeglądanie stron 170
1 2 ... 166 167 168 169 170 171 172 173 174 175 176 ... 611 612

Komentarze do niniejszej Instrukcji

Brak uwag