Voyager DP-263-S Podręcznik Użytkownika Strona 183

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 182
Sample Plate Types Supported
Voyager
Biospectrometry
Workstation User’s Guide 3-53
3
A .PLT file describes each position on the plate as discussed
in Table 3-4.
Table 3-4 .PLT File Parameters
Parameter Description
WellUnits
(optional)
Defines the units of the sample positions that are displayed in the
Manual Laser/Sample Positioning control page (see Figure 4-9 on
page 4-31). If you do not include the WellUnits parameter, the
software uses the default. Allowable values:
1 = Steps (default)
2 = Microns
WellShape
(optional)
Defines the shape of the sample positions that are displayed in the
Manual Laser/Sample Positioning control page. If you do not include
the WellShape parameter, the software uses the default. Allowable
values:
1 = Elliptical (default)
2 = Rectangular
If you are customizing a .PLT file, see “Shape” below.
WellWidth
(optional)
Defines the width of the sample positions that are displayed in the
Manual Laser/Sample Positioning control page. If you do not include
the WellWidth parameter the software uses the default (based on the
value selected for WellUnits). Default values:
Default Step value = 800
Default Micron value = 2,540
If you are customizing a .PLT file, see “Width” below.
WellHeight Defines the height of the sample positions that are displayed in the
Manual Laser/Sample Positioning control page. If you do not include
the WellHeight parameter the software uses the default (based on the
value selected for WellUnits). Default values:
Default Step value = 800
Default Micron value = 2,540
If you are customizing a .PLT file, see “Height” below.
Przeglądanie stron 182
1 2 ... 178 179 180 181 182 183 184 185 186 187 188 ... 611 612

Komentarze do niniejszej Instrukcji

Brak uwag